|
EPSRC Engineering Instrument PoolOffering FREE loans to the UK academic community |
|
| Home | Availability | Request form | Contacts | Gallery | Links | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| Home - availability and technical guides - Section K | Technical Guides: A - B - C - D - E - F - G - H - I - J - K | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Other InstrumentsDigital Image Correlation (DIC) is an optical method to measure deformation on an object's surface. The method tracks the grey value pattern in small neighbourhoods, called subsets, (indicated in red in the figure) during deformation. The surface of the subject must be covered with a speckle pattern. This can be applied e.g. by spray can.
The system is capable of measuring full-field, 3D displacement. 3D shape and the surface strain on complex geometries of many materials. Measurement fields from several mm² to >100m² can be covered by the system. Typical applications for 3D DIC are material testing and component testing:
The Instrument Pool has three DIC systems as described below. Limess Quasistatic 3D Digital Image Correlation System (3D DIC) Limess High Speed 3D Digital Image Correlation System (3D DIC) This system is supplied with four different calibration grids to suit different size subjects, two flood lights and tripods, a laptop loaded with camera control software and analysis software and associated cables and accessories. GAS CHROMATOGRAPH
The instrument has a clear ‘injector-column-detector’ configuration. The maximum inlet sample pressure is 2 bar at temperature up to 100°C via 1.6mm (1/16″) input port fitting. The carrier gas can be helium, hydrogen, nitrogen, argon, at pressure up to 10 bar, via 1/8″ Swagelok gas fitting. The instrument is fitted with 3 channels, one Flame Ionisation Detector (FID) channel, and two Thermal Conductivity Detector (TCD) channels. The detection limits are < 100 ppb (for the FID detector), and < 10 ppm (for the TCD detector). The columns can be changed to suit the application. Standard available columns can be easily installed using finger-tight couplings, including capillary narrow bore columns, wide-bore columns, and (micro) packed columns. Each column has a separate oven with temperature controlled in the range 30 to 175°C. The use of capillary columns for gas analysis reduces the analysis time, and a typical maximum runtime is 120 seconds. No additional flush or conditioning steps are necessary so the cycle time is equal to the runtime. All GC parameters for each channel, including valve switching, temperature and flow setting are programmable during the analysis using the ‘run-time-table’. A brochure is available from the manufacturer’s website http://www.gassite.com/compactgc FIBRE SPLICER
Special splicing functions are available, including splicing of various fibres such as dissimilar fibre, large diameter fibre, attenuation splicing and short cleaved length splicing. Features include:
Accessories provided:
A datasheet and technical specification can be downloaded from the manufacturer’s website: http://www.fujikura.co.uk/fibre_optics/products/splicers/index.html SENSL TIME CORRELATED SINGLE-PHOTON COUNTERTime-correlated single-photon counting (TCSPC) is a sensitive technique for recording low-intensity periodic light signals at extremely high time resolution. Specification and technical details
Accessories Optical rail, quartz lenses and XY micrometer mount for increased collection efficiency SKYSCAN 1174 CT SCANNER
The scanner uses an x-ray source with adjustable voltage from 20 - 50kV. The maximum samples size is 50mm in length, and from 5 – 30mm diameter, depending on magnification used. The x-ray image detector is a 1.3Mpixel CCD coupled to a scintillator via a zoom lens. The variable magnification (6-30 µm pixel size) and object positioning (50mm vertical travel) allows selection of the volume to be scanned. The scanner is controlled by a Dell T7400 desktop PC via two control cables and a FireWire (IEEE1394) input. A range of software is provided allowing fast volumetric reconstruction, 2D / 3D quantitative analysis, and realistic 3D visualization. Two filters are provided for use when scanning higher density specimens. Three sample stages are provided, and can be installed in place of the standard specimen holder. The stages are controlled by software independent of the main scanner control program.
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Degrees-of-Freedom |
6DOF |
| Number of Sensors | 1-16 |
| Update Rate | 240 Hz per sensor |
| Static Accuracy Position | 0.03in |
| Static Accuracy Orientation | 0.15° RMS |
| Latency | 3.5ms |
| Resolution Position at 30cm range | 0.0004cm |
| Resolution Orientation | 0.0012° |
| Extended Range Source | Up to 4.6 meters |
| Interface | RS-232 or USB (both included) |
The Polhemus Liberty LATUS (Large Area Tracking Untethered System) is a wireless system which consists of a 280/16 system electronics unit (SEU) along with 12 Wireless Markers and 16 signal receptors which are on a 18.3 m cable. Each receptor can track up to 12 markers in an area 2.44m diameter allowing large areas to be monitored by using multiple receptors. As this system uses low powered electromagnetic fields to make these measurements, care must be taken to avoid interference from the proximity of electrical systems and/or large metallic items such as steel building frameworks when using this system.
Degrees-of-Freedom |
6DOF |
| Number of Sensors | Wireless MARKERs 1-12 |
| Update Rate |
188Hz (1-8MARKERs) 94 Hz (9-12 MARKERs) |
| Static Accuracy Position |
0.254cm 1 MARKER Typical 2.54-7.62cm |
| Static Accuracy Orientation |
0.5° 1 MARKER Typical 1-3° |
| Latency | 5ms |
| Resolution Position at 30cm range | 0.0004cm |
| Resolution Orientation | 0.0012° |
| Range |
Range is dependable upon desired accuracy |
| Interface | RS-232 or USB (both included) |
Both systems are supplied with a Dell laptop PC complete with software ready installed to allow data capture and analysis.
At the heart of the ProLab is a well-proven quadrupole mass spectrometer analyzer. Mass spectrometry is particularly well suited to atmospheric monitoring due to the rapid speed of analysis and versatility of the technique. Furthermore, the certainty of identification of mass spectrometry is unmatched by other multicomponent techniques. This reduces the susceptibility to false alarms. For ionization, an enclosed ion source is used to achieve maximum sensitivity and resistance to contamination. Thoria filaments are used as they operate at low temperatures and give excellent operational life and ruggedness. A dual filament design is used to ensure maximum up time, allowing replacement at scheduled maintenance intervals.
The compact triple filter analyzer covers a mass range (1-300amu), which will enable the detection of most common samples. A triple filter design is employed to enhance performance in the 50-250 amu mass range and to further increase resistance to contamination. With the enclosed ion source and triple filter design maintenance requirements are reduced to one routine service every twelve months. The analyzer incorporates a proven dual detection system. This includes a secondary electron multiplier (SEM) detector that will be for detection of samples at ppm levels. A faraday detector is also supplied to monitor higher concentrations and to calibrate the SEM.
Typical applications
|
|
Specifications
| Application | Gas analysis at 1 atm |
| Analyser | Quadrupole mass spectrometer (triple mass filter) |
| Mass Range : | 1-300 amu |
| Ion Source : | Enclosed ion source |
| Detector : | Dual detector (Faraday/Channeltron multiplier) |
| Resolution : | 1 amu |
| Dynamic Range : | 7 decades (Faraday); 8 decades (Channeltron) |
| Min. Detectable : | <10 ppm (Faraday); <50 ppb (Channeltron) |
| Pumping : | Differential pumping with drag stage turbo pump and rotary backing pump |
| Inlet : | Heated capillary inlet with bypass |
| Bakeout : | To 200° C |
| Electronics Engine : | Smart IQ+ |
| Gauges : | Bakeable Cold-Cathode gauge |
| Filaments : | Two thoriated iridium filaments, for high sensitivity with low backgrounds |
| STFC Contacts | STFC News | How to Find STFC | STFC Publications | Vacancies | Site Map | Top of the Page |
|
Freedom of Information Privacy Statement Disclaimer |
|
|